Impedance Eye Recording IER: A New Glaucoma Diagnostic Tool?

Letter to the Editor

J Ophthalmol & Vis Sci. 2021; 6(3): 1055.

Impedance Eye Recording IER: A New Glaucoma Diagnostic Tool?

Grounauer PA¹*, Moreno MV² and Houriet C³

1University Eye Clinic, FMH ophthalmologist, Lausanne, Switzerland

2Biomedical Engineering, CEO Runsys, Lyon, France

3HES engineer, CEO Fabrinal SA, La Chaux-de-Fonds, Switzerland

*Corresponding author: Grounauer PA, University Eye Clinic, FMH ophthalmologist, Lausanne, Switzerland; Email: [email protected]

Received: August 14, 2021; Accepted: September 11, 2021; Published: September 18, 2021

Letter to the Editor

Described in 1941 by two New Yorker physicist brothers, KS Cole et RH Cole, who propose and publish together a new mathematical equation plotting polymers’ electrical reactance/resistance correlation, Rea/Res, named the Cole-Cole curve [1]. Since then, this method is used to know how sinusoïdal current goes through living tissues, be they vegetable or animal. Applied on the human body it is included in weighing machines to adapt sports training, for example. In medicine to evaluate many organs like lungs, breast [2] or skin tumours, heart and brain or to separate normal from tumoral cells. In Clinical Ophthalmology to our knowledge, they are not yet similar papers. At the present time PubMed, key words are for first eye impedance or eye and clinical impedance spectroscopy, in August 2021, 22 references without strict IER correlations and then none. Closest publications are Fukuda [3] to quantify corneal defects, Jürgens [4] about in vitro eye impedance and CA Gutierrez [5] about lens hydratation.

In the University Eye Clinic electrophysiology lab, Lausanne Switzerland, together with MVM [6] and Fabrinal [7] we have modified the ERG-Jet (electroretinogram) contact lens from an electric retinal captor to an emitting electrical injector, thus describing and creating sure IER scientific basic data to obtain acceptable impedance ocular signal/noise recordings. Our first results show this method is easy to use, quick and without discomfort. Furthemore we found Rea/Res age-dependent [8]. Now we are in the process of collecting normal IER values on healthy young people to evaluate sensitivity then specificity. In the recent Springer paper [9] we describe a phantom glaucoma model to simulate extra/intra cell water concentrations in the ocular anterior chamber. Therefore, we found this pilot study a good predictive model. That is the reason why we presume that this relatively new IER providing Cole-Cole ocular curve can probably be a useful complementary tool to understand glaucomas and/or other ophthalmic diseases where extra/intra cell water concentrations are abnormal.

References

  1. Cole KS, Cole RH. Dispersion and Absorption in Dielectrics I. Alternating Current. Journal of Chemical Physics. 1941; 9: 341-351.
  2. Moreno M-V, Herrera E. Evaluation of the Feasability of a Smart Bra to Detect Breast Cancer in Young Adults. Sensors. 2019; 19: 5491
  3. Fukuda M, Sasaki H. In vivo measurement of human corneal impedance value. Cornea. J. 2016; 35: 10.
  4. Jürgens I, Rosellz J, RIu PJ. Electrical impedance tomography of the eye: In vitro measurements of the cornea and the lens. Physiol. Meas. 1996; 17: A187-A195.
  5. C Gutierrez. Impedance-base measurement of eye-mountable lens hydratation. Pat number WO2018156772A1, Verily life science. 2018.
  6. Runsys.eu
  7. Fabrinal.ch
  8. Moreno M-V, Coux MN, Clarion A, Grounauer PA. Electrical evolution of aqueous humor with ageing. ICEBI. 2016; 112: 16.
  9. Moreno M-V, Houriet C, Grounauer PA. Ocular Phantom-Based Feasability Study of an Early Diagnosis Device for Glaucoma. Sensors. 2021; 21: 579.

Download PDF

Citation: Grounauer PA, Moreno MV and Houriet C. Impedance Eye Recording IER: A New Glaucoma Diagnostic Tool?. J Ophthalmol & Vis Sci. 2021; 6(3): 1055.