Investigation of Intraobserver Repeatability and Interobserver Reproducibility of Corneal Indices Obtained Using Three Different Topographers

Research Article

J Ophthalmol & Vis Sci. 2022; 7(2): 1070.

Investigation of Intraobserver Repeatability and Interobserver Reproducibility of Corneal Indices Obtained Using Three Different Topographers

Al Saqr A*, AL Dleam A, Almutairi S, Alharbi A and Abu Sharha A

Department of Optometry, College of Applied Medical Sciences, King Saud University, Saudi Arabia

*Corresponding author: Al Saqr A, Department of Optometry, King Saud University, College of Applied Medical Sciences, Saudi Arabia

Received: August 09, 2022; Accepted: September 05, 2022; Published: September 12, 2022

Abstract

Purpose: Investigating intraobserver and interobservers repeatability, reproducibility and limits of agreement of anterior corneal indices using three different topographers.

Methods: Forty five participants were recruited, involved two examiners and two sessions. They were assessed in random order with EyeSys Vista, Medmont E-300 and manual keratometer. The corneal indices were horizontal keratometer, vertical keratometer, eccentricity and inferior superior index. Repeatability and reproducibility of corneal power measurements were assessed based on the intersession, intrasession, Intraclass Correlation Coefficient (ICC), Spearman’s correlations and agreement was evaluated by Bland Altman test (95% Limits Of Agreement, LoA).

Result: All devices showed high repeatability (p < 0.01) and reproducibility (ICC>0.75) of the corneal indices. Multivariate analysis of variance showed nonsignificant findings between visits and examiners (p > 0.05), while there was a significant difference between the devices among keratometric readings (p < 0.0001). The differences between the EyeSys, Medmont and manual keratometer were significant (p < 0.01). Moderate relationship was found between the keratometric readings with the three devices (r = 0.60, p < 0.0001). The 95% LoA for EyeSys and Medmont of the keratometric readings were larger than those for the Medmont and manual keratometer (±0.3 mm). The eccentricity and inferior superior index measurements were not significantly different between the EyeSys and Medmont (p > 0.01).

Conclusion: All devices exhibited excellent repeatability and reproducibility. The EyeSys is portable, repeatable, and reliable and have stable interobservers variability. Although EyeSys would not be used interchangeably with the other two devices due to lower agreement, corrected linear calculation was suggested to manage the differences between the EyeSys and the two devices.

Keywords: Corneal curvature; Corneal topography; Keratometry; EyeSys vista; Medmont

Abbreviations

IOL: Intraocular Lens; SD: Standard Deviation; D: Diopters; VA: Visual Acuity; OSDI: Ocular Surface Disease Index; Kh: Horizontal K-Reading; Kv: Vertical K-Reading; Ecc: Corneal Eccentricity; IS: Inferior Superior Index; ICC: Intraclass Correlation Coefficient; Loa: Limits of Agreement; IQR: Interqurtial Range; MANOVA: Multivariate Analysis Of Variance

Introduction

The cornea provides two-thirds of the total optical power of the eye. Determining corneal curvature and other corneal indices in a reliable and accurate method is vital in clinical procedures and research setting. Measuring corneal curvature is essential in assessing contact lenses, keratoconus signs, calculating intraocular lens (IOL) and refractive surgery procedure [1-4]. Numerous devices are commercially available for the assessment of corneal parameters. Several studies have investigated the accuracy and reliability of those measurements obtained using different devices [5-10]. Studies to evaluate the interchangeability of measurements obtained from the different devices are essential to confirm if they can provide comparable data.

One of the oldest and traditional device to measure the corneal curvature is the manual keratometer which widely considered to be a gold standard device [11]. In the last decades several automated computerized topographershas been developed and investigated. Two of those are placido disk–based corneal topographers, Medmont E300 (Medmont Pty. Ltd., Victoria, Australia) and EyeSys Vista (EyeSys INC., Texas, USA)). The main aim of this study is to assess the accuracy, reliability and reproducibility between two examinersover two visits with those two devices as well as the manual keratometer.

The desktop Medmont E-300 is a placido disk-based topography, has 32 placido rings, which utilize arc-step reconstruction algorithm and incorporates a range finder [12,13]. It measures the distance from the corneal apex to the device camera and then automatically takes images. For each image captured a score out of 100 based on focus, movement and centering was given and the image scores >75 was considered of good quality. The Medmont measures the central 3 mm diameter field of the corneal radius of curvature in both the flat and steep meridians.

Silmilar to the Medmont E-300, the portable EyeSys Vista is a placido-based corneal topography that has 25 placido rings and its measurement can be taken within 3, 5 and 7mm diameter field of the corneal. The topography map then sent to the iTrace system with integrated software [5]. The EyeSys Vista advantageous features over the other similar devices are portability and freedom of use outside the conventional clinic or lab. In this study, the 3 mm diameter field readings were selected in order to match areas of measurement in the other two devices.

As the Medmont E-300 and the EyeSys Vista devices have the same optical principles behind them, therefore it is likely to have minimal differences when assessing corneal measurements. Prior study investigated the intrasession and intersession of those two devices among other 6 devices [5]. Although, a single examiner took all measurements in the two sessions, that procedure would not show variability of the devices between examiners.

The main aim of the present study is to prospectively determine the variability (between examiners), repeatability (intrasession) and reproducibility (intersession) of corneal indices using the 3devices mentioned earlier and to investigate differences in keratometric and corneal asphericity for each of the different devices in order to check the interchangeability between those devices in healthy eyes.

Materials and Methods

This study obtained an institutional ethical board approval. The aims of the study and any potential consequences were fully described for all participants, and self-informed written consent was collected. This study adhered to the tenets of the Helsinki Declaration at all stages.

The study is a cross sectional in design and recruited participants aged 18-45 years. The study aimed to comprehensively compare three instruments: the Vista system (Eye Sys Vision, Houston, TX, USA), the Medmont E-300(Medmont Pty. Ltd., Victoria, Australia) and the Bausch and Lomb manual keratometer (Bausch and Lomb, Quebec, Canada).

Sample Size

The sample size was calculated with a PS Power and Sample Size Calculation Software (version 3.1.6, Vanderbilt University, USA). The pooled Standard Deviation (SD) of the differences in corneal radius of curvature between several instruments was found in previous study to be approximately 0.10 diopters (D) [14]. The criteria entered was two-sided test of a level of significance (a) = 0.05 and a power of 99%, a sample size of 21 pairs (42 participants) was suggested to be the required sample in order to detect a difference of 0.10 D between different instruments.

Procedure

The recruited participants were excluded if they have 1) distance Visual Acuity (VA) worse than 20/25 to avoid any miss-fixation, 2) dryness based on the ocular surface disease index (OSDI) questionnaire (with a cut off of 13 as suggested by Schiffman et al.) [15], 3) contact lens wearers, 4) have any corneal opacities, corneal scars, cataracts, or any previous ocular surgery.

The refraction assessment was conducted using the Nidek OPDScan III (Nidek Technologies, Gamagori, Japan) in accordance with the manufacturer’s directions by means of the auto-tracking and auto-shot functions. Briefly, the participants were instructed to fixate on the image inside the device, which has an auto fogging feature to relax the accommodation. The device automatically takes for each eye three readings and the representative reading signposted by parentheses which was then selected [16].

Each participant was tested with the three devices in random order on two visits by two examiners. Both examiners were masked to each other’s record and both were trained and experienced in using the 3 topographers. The participant was tested at the same location, between 9:00 AM to 4:00 PM to minimize variations in the results. In each visit, the examiners collected the measurements with each instrument for all participants according to the manufacturers’ instructions.

The protocol of measuring the anterior corneal indices followed the same procedure of a previous study [5]. Briefly, each examiner asked each participants to blink completely just before each scan. Then asked to sit back after each measurement. The instrument was readjusted before each measurement. The measurements with the 3 topographers were in continual procedure, and without any significant time intervals. The repeated measurements were scheduled within one week at approximately the same time as the first visit, using the same protocol.

This study compared several corneal shape descriptors between the Medmont E300 and the vista system while the Keratometer focuses only on K-readings. For each instrument: horizontal and vertical K-reading (Kh and Kv), eccentricity (Ecc), eccentricity indicates the departure of the peripheral curvature of the cornea from the apical radius and refers to degree of asphericity (the average normal corneas would have an eccentricity of about 0.55); [17] Inferior Superior index (IS), which refers to inferior-superior dioptric asymmetry an indicative risk factor keratoconus[18].These descriptors are the key determinates of the corneal power, shape, and integrity [17,19,20].

Data Analysis

The data analysis was conducted using IBM SPSS Statistics (IBM Corp., Armonk, N.Y., USA). The data was not normally distributed (Kolmogorov-Smirnov test, p<0.05), the median± Interqurtial Range (IQR) were used to report the data. The differences between examiners, devices and visits were considered statistically significance when the p value was <0.01, in order to consider the Bonferroni correction.

It has been suggested that there is eventual correlation existing between right and left eyes of a single patient. [21,22] The relationship between the right and left eye measurements in the 3 instruments were investigated, and very strong and significant relationship was found (r > 0.80, p <0.05). Therefore, to follow previous studies procedures and to avoid any bias the outcome of right eye from each participant with the 3 instruments was selected in the further analysis.

Comparison within Instruments

Intrasession repeatability was assessed with Intraclass Correlation Coefficient (ICC) and their respective 95% CI. The ICC measure the consistency for data sets of repeated measurements. According to previous suggestion the closer the ICC is to 1, the better the measurement consistency [5]. Specifically, the ICC score of >0.75 indicates “excellent,” 0.40 to 0.75 “fair to good,” and <0.40 “poor” reliability [23]. Additionally, Wilcoxon signed rank test was used to investigate the difference within each examiners, while the Mann- Whitney test was used to investigate the difference between the two examiners and topogrpahers.

Bland-Altman analysis (95% Limits Of Agreement (LoA)) were also used to express the extent of agreement between test-retest outcomes [24]. In this test the difference of the paired intra and intersession measurements is plotted against the corresponding average. It has been recommended that 95% of the data points lies within the mean ± 1.96 SD of the differences for the intrasession and intersession measurements, which corresponds to the 95% confidence interval. A narrower 95% limit of agreement suggests better agreement between instruments and/or examiners.

Inter-session reproducibility was investigated by exploring both the ICC for the repeated measurements obtained during the first and second visit. And the Bland-Altman analysis was conducted to determine the extent of agreement between measurements during the first and second visit.

Comparison between Instruments

Multivariate Analysis Of Variance (MANOVA) was conducted while also taking into account the Bonferroni correction to show pairs that were significantly different (p < 0.01). Bland-Altman analysis was further applied to investigate the 95% LoA between instruments. Finally, the Spearman’s test was conducted to explore the relationship between the three topographers’ scores as well as the corrected linear calculation in order to suggest how to correct those findings to match each other perfectly.

Result

This study recruited 45 participants, each one of them tested twice with two independent examiners. The participants’ age ranged from 22 to 45 years with a median of 21 ± 4.00 years. The spherical component was Plano ± 1.25 (ranges +2.00 to -7.00); while the spherical equivalent was Plano ± 1.00 (ranges 1.75 to -7.50). The astigmatism component was converted into vector representation, J0 (cylinder at 0-degree meridian) and J45 (cylinder at 45-degree meridian). The median of vector J0 was 00 ± 0.80 (ranged from -1.50 to 1.00), and vector J45 was 00± 0.12 (ranged from -1.50 to 1.50).

The summary of the measured Kh, Kv, eccentricity and the IS are listed in (Table 1). In general, it was observed that the EyeSys Vista provided flatter K-readings than those observed with the other two devices (Table 1). Both the Medmont E300 and EyeSys Vista provided similar eccentricity readings and indicated that the participants had slightly higher score than the suggested average normal corneas. Although, both devices provided similar IS scores, the Vista device had a larger interquartile range than the Medmont E300 device.