CT Image Quality and Dose Reduction Management

Research Article

Austin J Radiol 2023; 10(1): 1207.

CT Image Quality and Dose Reduction Management

Diab HM1*, Kameesy SUE2, Mostafa F3 and Abou-Elenein HS4

1Nuclear and Radiological Research Safety Center, Egyptian Atomic Energy Authority (EAEA), Egypt

2Department of Physics, Faculty of Science, Ain Shams University, Egypt

3Ministry of Health, Egypt

4Department of Radiotherapy, Children Cancer Hospital, Egypt

*Corresponding author: HM Diab Nuclear and Radiological Research Safety Center, Egyptian Atomic Energy Authority (EAEA), Egypt

Received: November 24, 2022; Accepted: January 03, 2022; Published: January 09, 2023

Abstract

Medical imaging is characterized by tremendous and undeniable benefits for patients in recent health care. Combining two imaging methodologies into a single examination, such as SPECT/ CT and PET/CT, so-called hybrid imaging, is also increased. Many factors may affect the image quality. Different CT protocols are designed to perform high-quality examination at low radiation dose. The present study is focused on the assessment of the performance of image quality for fifteen CT scanners with different scan mode (from 2slice to 320 slice) at 80, 100 and 120 kV with tube current of 50–250 mA to optimize the radiation dose. Also, the uniformity was examined as part of the quality control tests. The results showed that, the 320 slice CT scanner delivered the minimal radiation dose compared with the 16 and 64 slice CT scanners which mostly applied in Egyptian centers (governmental and private). Both 16 and 64 slice CT scanners produce acceptable image quality but with associated high absorbed dose. These results give a further support to the priority of the 320 slice CT scanner to be applied. Field uniformity measurements were adopted and the difference in the calculated values for different protocols was precisely evaluated and discussed.

Keywords: CT image quality; Image noise; Contrast; Resolution

Introduction

The objectives in CT development have changed from increasing the number of slices to focusing on improvements in X-ray tube performance, detector efficiency and data processing [1]. After the introduction of Multi-Slice CT (MSCT) in 1997 [2], the number of slices acquired per rotation has rapidly increased from 2 up to 4, 8, 16, 32, 40, 64, 128 and 320 [3]. The primary advantage of MSCT is improved temporal (<250 ms), spatial resolution (<0.5 mm) and smaller scan times [4]. Since 2006, a new scanner technology using two X-ray sources and two detectors simultaneously, dual source CT, have been settled up [5]. The technology has provided further improvements in scan speed and temporal resolution (0.28 s rotation time and 75 ms temporal resolution). By utilizing Dual Energy CT (DECT), using either dual source or kV-switching, advanced post-processing and visualization, new clinical applications have been found. The advantage with DECT is that the properties of X-ray attenuation change at different energies, which are used to differentiate materials, including iodine, calcium, and uric acid crystals. In recent years, iterative reconstruction methods have been introduced that provide great potential for improving image quality and reduced radiation doses [6].

Image quality is essential and can be used as a quality assurance/ control indicator for imaging system performance and also for optimizing patient radiation dose during X-ray practice. The challenge is how to establish protocols for specific diagnostic with sufficient image quality and lowest effective dose to patient.

The most known detector configuration today is from 64 to 320 detector rows. Every CT slice is subdivided into a matrix (e.g., 512×512 or 1024×1024). Each element (voxel) has been traversed by X-ray photons, and the transmitted radiation is detected by the detector. The attenuation of a monochromatic narrow X-ray beam going through a homogeneous material is described by the following equation:

I = I e-μx (1)

where I is the number of photons behind the object, I0 is the number of photons at the same point in the absence of the object, x is the object thickness, and μ is the linear attenuation coefficient of the material for the photon energy used. The linear attenuation coefficient μ is strongly dependent on the photon energy and therefore only of limited use for characterizing the radiation attenuation capacity of an object in CT. The attenuation value in CT (Hounsfield unit; HU) is the scaled difference of the linear attenuation coefficient of the investigated object from the linear attenuation coefficient of water. Water is used as the reference material:

HU = k*((μobject - μwater) /(μwater - μair)) (2)

where k is 1000 and μwater and μair are the linear attenuation coefficients of water and air. Specific attenuation values are assigned to each individual voxel. The reconstructed image consists of a matrix of picture elements or pixels. Each pixel is assigned a numerical value (HU) which is the average of all attenuation values within the voxel. This scale assigns for water an attenuation value of 0 HU and for air 1000 HU. Each number represents a shade of grey with - 1000 HU (black) to +1000 HU (white). By using so-called windowing technique (window widths and levels), certain types of tissues can be viewed in more detail [7].

There are several metrics describing different aspects of image quality in CT. Noise, uniformity high-contrast spatial resolution and low-contrast spatial resolution [8]. Several methods are available to optimize and minimize the radiation dose in CT [9,10]. The scanning parameters should be optimized for each specific examination and special efforts should be made with pediatric CT protocols [11]. A number of scanning parameters influence patient radiation dose and image quality: tube current, tube voltage, filtration, collimation, reconstruction method, reconstruction filter, slice thickness, pitch, and scanning length [12]. The operator can monitor most of these parameters and modify them to obtain the necessary image quality with a minimal absorbed dose to the patient. A simple relationship exists between the tube load (the product of tube current and exposure time per rotation, mAs) and radiation dose to the patient. A 50% reduction in tube load reduces the radiation dose by one half its values but also increases the noise level by a factor of v2. An adequate mAs level can be determined using dose reduction simulation software [13]. The software adds artificial noise to the CT raw data to simulate a scan acquired with lower dose (mAs). The tube voltage determines the energy of the emitted photons from the X-ray tube; consequently, a variation in tube voltage changes the radiation dose and image quality. Reduction in tube voltage results in reduced radiation dose when all other parameters are held constant. This will increase the image noise and cause contrast changes. Several studies have demonstrated an ability to affect radiation dose and image quality by using a lower tube voltage to optimize the Contrastto- Noise Ratio (CNR) and minimize the absorbed dose [14,15].

In CT, Automatic Exposure Control (AEC) is automatically modulated the current of tube in the plane of x-y (angular modulation) along the scanning direction (z-axis; longitudinal modulation) or both (combined modulation) [16]. The modulation is done according to the size of the patient, shape, and the attenuation of the scanned body parts. The system adjusts the current of the tube to obtain the pre-determined image quality indicated by the operator with improved radiation efficiency. The adaptation of the tube current is based on attenuation data from the localization radiograph and attenuation profiles or feedback from online measurements. AEC systems have several benefits: better control of the absorbed dose to the patient, improved consistency of image quality, reduction of certain artifacts related to the image, and small load on the X-ray tube, which increases its lifetime [17]. All modern CT systems are characterized by AEC systems that modulate current of the tube in three dimensions. Each of them has different functional properties. However, the main role is to adapt the needed image quality and radiation dose in a reproducible method by managing the tube current to the patient’s size, shape, and attenuation.

This study was conducted with the aim to evaluate the performance of CT scanners in governmental and private hospitals at various kilovolt and tube current setting. Image quality was assessed using a CT image quality phantom. The evaluated Image quality parameters are CT Image noise, uniformity and Contrast.

Materials and Methods

CT Scanners

The data used in the present work were collected from thirteen departments of radiology (private and governmental sectors) as shown in Table 1. The most wide spread procedures performed in the radiology department are head and abdomen.